1. INTRODUCTION
The Field Logger is versatile data acquisition equipment for process variables,
capable of on-line or stand alone operation. When loaded with a Real Time Clock and
non-volatile memory, it can register up to 128k readings. Its universal input channels
accept a large variety of sensors and standard signals as listed in table Field Logger
Input Signals, below.
A Configuration program (Windows 95, 98 or NT) is provided for an easy
configuration of all the internal parameters.
2. PRESENTATION
Technical Parameters
Mains: 85 to 250 V
AC
(50/60Hz) or 100 to 250 V
DC
, or 24 V
AC
or V
DC
(optional)
Power consumption: 2VA
All sensors factory calibrated and software linearized.
Internal cold junction compensation for thermocouples, 2 or 3 Wire Pt100.
Pt100 input: 3 wire configuration with cable length compensation.
Excitation current: 170A
Internal resolution: 20000 counts
Linearity: Better than 0.05% of FS
Accuracy
Thermocouple J, K, T, E and N: 0.2% of range ±1°C.
Thermocouple R, S and B: 0.25% of range ±3°C.
Pt100: 0.2% of range
Linear voltage or current: 0.2% of maximum range
RS485 interface (MODBUS RTU)
Ambient temperature: 0 to 55
o
C (32 to 130
o
F)
ABS case
Dimensions: 1059060mm
Weight: 210 grams
Alarms: 2 SPST-NA - 3A / 250V relays